Output 1
Ranges: Range 1 - 0V to 35V at 1mA to 3A; Range 2 - 0V to 16V at 1mA to 6A
Operating Mode: Constant voltage or constant current with automatic cross-over
and mode indication.
Voltage Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 1mV.
Current Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 1mA.
Setting Accuracy: Voltage - 0.05% of reading ± 3mV;
Current - 0.3% of reading ± 3mA to 3A, 0.5% of reading ± 3mA to 6A
Setting Stores: Up to 50 set-ups can be saved and recalled via the keyboard
(or the digital interfaces on MX100TP).
Load regulation: <0.01% +5mV (CV mode) for any load change using remote sense.
Line regulation: <0.01% +5mV (CV mode) for a 10% line voltage change.
Ripple & Noise: Typically <0.5mV rms, <5mV pk-pk, 1mV rms max. (20MHz bandwidth).
Transient Response: <100us to within 50mV of set level for 5% to 95% load change.
Over Voltage Trip: Settable 1V to 40V in 0.1V steps
Over Current Trip: Settable 0.1A to 7A in 0.01A steps
Sensing: Selectable local or remote sensing.
Output 2
Ranges: Range 1 - 0V to 35V at 1mA to 3A; Range 2 - 0V to 16V at 1mA to 6A
Range 3 - 0V to 35V at 1mA to 6A* (available when output 3 is disabled).
Operating Mode: Constant voltage or constant current with automatic cross-over
and mode indication.
Voltage Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 10mV.
Current Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 10mA.
Setting Accuracy: Voltage - 0.1% of reading ± 10mV;
Current - 0.3% of reading ± 3mA to 3A, 0.5% of reading ± 3mA to 6A
Setting Stores: Up to 50 set-ups can be saved and recalled via the keyboard
(or the digital interfaces on MX100TP).
Load regulation: <0.01% +5mV (CV mode) for any load change using remote sense.
Line regulation: <0.01% +5mV (CV mode) for a 10% line voltage change.
Ripple & Noise: Typically <0.5mV rms, <5mV pk-pk, 1mV rms max. (20MHz bandwidth).
Transient Response: <100us to within 50mV of set level for 5% to 95% load change.
Over Voltage Trip: Settable 1V to 40V in 0.1V steps
Over Current Trip: Settable 0.1A to 7A in 0.01A steps
Sensing: Selectable local or remote sensing.
Output 3
Ranges: Range 1 - 0V to 35V at 1mA to 3A; Range 2 - 0V to 70V at 1mA to 1.5A
Range 3 - 0V to 70V at 1mA to 3A* (available when output 2 is disabled).
Operating Mode: Constant voltage or constant current with automatic cross-over
and mode indication.
Voltage Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 10mV.
Current Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 1mA.
Setting Accuracy: Voltage - 0.1% of reading ± 10mV;
Current - 0.3% of reading ± 3mA
Setting Stores: Up to 50 set-ups can be saved and recalled via the keyboard
(or the digital interfaces on MX100TP).
Load regulation: <0.01% +5mV (CV mode) for any load change using remote sense.
Line regulation: <0.01% +5mV (CV mode) for a 10% line voltage change.
Ripple & Noise: Typically <0.5mV rms, <5mV pk-pk, 1mV rms max. (20MHz bandwidth).
70V range - typically <1mV rms, <10mV pk-pk, 1.5mV rms max.
Transient Response: <100us to within 50mV of set level for 5% to 95% load change.
Over Voltage Trip: Settable 1V to 80V in 0.1V steps
Over Current Trip: Settable 0.1A to 3.5A in 0.01A steps
Sensing: Selectable local or remote sensing.
Output Protection
External Voltage: Output will withstand forward voltages of up to 50V (O/Ps 1 and 2) or
80V (O/P 3). Reverse protection by diode clamp, 3A max..
Fault Trip: The output will be shut down if a trip conditions listed below occurs.
OVP or OCP Exceeding over-voltage or over-current settings for the output.
Over Temperature: Monitors internal temperature rise to protect against excess ambient
temperature or blocked ventilation slots.
Connection
Output Terminals: Universal 4mm safety binding posts on 19mm (0·75”) spacing at front.
Screw terminals at rear (MX100TP only).
Terminals can accept fixed shroud 4mm plugs, standard 4mm plugs, fork terminals and bare wires.
Sense Terminals: Sprung loaded screw-less terminals at front.
Screw terminals at rear (MX100TP only).