| Specifications 1-Year | 
| Low pressure sensor | Range | Resolution | Accuracy | 
| -14 psi to +16 psi, -0.97 bar to 1.1 bar
 | 0.001 psi, 0.0001 bar
 | 0.025 % of full scale | 
| High pressure sensor | -12 psi to +100 psi, -0.83 bar to 6.9 bar
 | 0.01 psi, 0.0001 bar
 | 0.025 % of full scale | 
| General Specifications | 
| Instrument setup recall | Last used on power-up | 
| Environmental | Operating temperature -10 °C to +50 °C Storage temperature (without batteries) -20 °C to +60 °C
 | 
| Altitude | 2,000 meters | 
| Ingress protection | IP-40 | 
| Power requirements | 6.0 v dc | 
| Battery life | >35 hours, typical usage | 
| Batteries required | (4) Standard AA cells | 
| Dimentions (HxWxD) | 7.9" X 4.3" X 2.3" (20 cm X 11 cm X 5.8 cm) | 
| Weight | 03539 kg (1 lb 3 oz) | 
| Connectors/ports pressure | Two, 1/8 in NPT, 700P Series Pressure Module; RTD Probe | 
| Electrical and Temperature Measurement 1-Year | 
| Function | Range | Resolution | Accuracy | 
| mA measure | 0 to 24 mA | 0.001 mA | ±0.015% of rdg ±0.002mA
 | 
| VDC measure | 0 to 30 VDC | 0.001V | ±0.015% of rdg ±0.002V
 | 
| *Temperature measure (RTD/ Ohms) | -40 °C to 150 °C (-40 °F to 302 °F)
 | 0.01 °C, 0.01 °F | ±0.015% of rdg ±0.02 ohms
 ±0.1°C (±0.2°F)
 ±0.25 degC (±0.45
 degF) combined uncertainty when using 720 RTD probe accessory
 | 
| Loop power supply | 24V | N/A | 24 mA @ 24V | 
| *Temperature measurement requires optional Fluke-720RTD Pt-100 RTD probe available as an accessory | 
| Temperature Effect | 
| No effect on accuracy on all functions from 15 °C to 35 °C | 
| Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C | 
| Available Pressure Engineering Units | 
| psi, bar, mbar, kPa, kgcm2, cmH2O@4°C, cmH2O@20°C, mH2O@4°C, mH2O@20°C, inH2O@4°C, inH2O@20°C, inH2O@60°F, mmHg@0°C, inHg@0°C, ftH2O@4°C, ftH2O@20°C, ftH2O@60°F |